Vyaire Medical recalls AirLife Adult Manual Resuscitator
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
- Products
-
- AirLife Adult Manual Resuscitator
- Company
- Vyaire Medical
- Where it was sold
-
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Nationwide
- Units affected
- 106,656 units
- About this recall
- AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,