Zyno Medical LLC recalls
Zyno Medical LLC is named on 6 recalls in Recall Relay's records since November 2023, from FDA (9 recalled products: FDA lists each product as its own recall); the most recent was May 7, 2025.
All 9 recalled products, newest first.
-
Class Iofficial
Zyno Medical LLC recalls Z-800F Infusion System. Intended to provide intravenous infusion of parenteral fluidsUnreleased software versions were installed on distributed devices without verification or validation.
-
Class Iofficial
Zyno Medical LLC recalls Z-800W Infusion System. Intended to provide intravenous infusion of parenteral fluidsUnreleased software versions were installed on distributed devices without verification or validation.
-
Class Iofficial
Zyno Medical LLC recalls Z-800WF Infusion System. Intended to provide intravenous infusion of parenteral fluidsUnreleased software versions were installed on distributed devices without verification or validation.
-
Class Iofficial
Zyno Medical LLC recalls Z-800 Infusion System. Intended to provide intravenous infusion of parenteral fluidsUnreleased software versions were installed on distributed devices without verification or validation.
-
Class IIofficial
Zyno Medical LLC recalls Z-800 Infusion System. Model/Catalog Number: Z-800Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
-
Class IIIofficial
Zyno Medical LLC recalls Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion systFor the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
-
Class Iofficial
Zyno Medical LLC recalls Z-800 Infusion SystemThere is a defect in the air-in-line software algorithm.
-
Class IIofficial
Zyno Medical LLC recalls Zyno Medical Z800Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
-
Class IIofficial
Zyno Medical LLC recalls Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluidsWhen utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.