Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Zyno Medical LLC recalls Zyno Medical Z800

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
Products
  • Zyno Medical Z800
Where it was sold

US Nationwide distribution.

Nationwide

Units affected
88 units
About this recall
Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps