Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Qualigen Inc recalls FastPack TSH Calibrator Kit

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
Products
  • FastPack TSH Calibrator Kit
Where it was sold

US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland

States: Alaska, Alabama, Georgia, Iowa, Kansas, Massachusetts, Minnesota, Mississippi, Montana, North Carolina, Nebraska, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, South Dakota, Texas, Utah, Vermont, Washington, Wisconsin, West Virginia, Wyoming

Barcodes
00816467020099
Units affected
60
About this recall
FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer