Qualigen Inc recalls The FastPack IP Sex Hormone Binding Globulin Immunoassay
FDA Class III
FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).
- Hazard
- Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
- Products
-
- The FastPack IP Sex Hormone Binding Globulin Immunoassay
- Company
- Qualigen Inc
- Where it was sold
-
Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.
States: Arkansas, California, Colorado, Florida, Illinois, Indiana, Massachusetts, Missouri, North Carolina, New York, Ohio, Oklahoma, Tennessee, Texas, Virginia
- Barcodes
20816467020345- Units affected
- 426
- About this recall
- The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.