All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Qualigen Inc recalls The FastPack IP Sex Hormone Binding Globulin Immunoassay

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
Products
  • The FastPack IP Sex Hormone Binding Globulin Immunoassay
Where it was sold

Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.

States: Arkansas, California, Colorado, Florida, Illinois, Indiana, Massachusetts, Missouri, North Carolina, New York, Ohio, Oklahoma, Tennessee, Texas, Virginia

Barcodes
20816467020345
Units affected
426
About this recall
The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.