Medtronic Navigation, Inc. recalls StealthStation S8 App versions 1.3.0 and 1.3.2
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to two software anomalies that can occur within a spine procedure. Two issues are: 1. Navigation Orientation - when lateral orientation radiographic images are taken, information may appear flipped when utilizing the system (180 degree flipped). 2. Pop-up Message - when using CT+Fluouroscopy images with a slice spacing thickness greater than 2mm, the system does not display a pop-up message indicating the slice spacing is not optimal.
- Products
-
- StealthStation S8 App versions 1.3.0 and 1.3.2
- Company
- Medtronic Navigation, Inc.
- Where it was sold
-
Worldwide - U.S Nationwide distribution including in the states of AK, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, TN, TX, UT, VA, WA, WI and WV. The countries of Argentina, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, Brazil, BULGARIA, CANADA, CHILE, China, Colombia, Costa Rica, CROATIA, CYPRUS, DENMARK, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, HUNGARY, ICELAND, India, ISRAEL, ITALY, Japan, Jordan, KAZAKHSTAN, KENYA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, Malaysia, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NICARAGUA, OMAN, PAKISTAN, PANAMA, Philippines, POLAND, PUERTO RICO, QATAR, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, Singapore, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, Turkey, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM.
Nationwide
- Barcodes
0076300045698600763000368494- Units affected
- 1,801 systems
- About this recall
- StealthStation S8 App versions 1.3.0 and 1.3.2, (Model # 9735762) The StealthStation System, with StealthStation Spine software, is intended as an aid for precisely locating anatomical structures in either open or percutaneous neurosurgical and orthopedic procedures.