Medtronic Navigation, Inc. recalls Medtronic Biopsy Needle Kit
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
- Products
-
- Medtronic Biopsy Needle Kit
- Company
- Medtronic Navigation, Inc.
- Where it was sold
-
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
States: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming
- Barcodes
0064316970294300643169030121- Units affected
- 3746 units
- About this recall
- Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly