Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls Model Number U225 VISIONIST CRT-P EL

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Products
  • Model Number U225 VISIONIST CRT-P EL
Where it was sold

Worldwide

Nationwide

Barcodes
00802526559426008025265594330080252657711600802526578809
Units affected
12,326 units
About this recall
Model Number U225 VISIONIST CRT-P EL