Boston Scientific Corporation recalls
Boston Scientific Corporation is named on 50 recalls in Recall Relay's records since November 2023, from FDA (266 recalled products: FDA lists each product as its own recall); the most recent was September 10, 2026.
The newest 200 of 266 recalled products, newest first.
-
Class IIofficial
Boston Scientific Corporation recalls NaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.
-
Class Iofficial
Boston Scientific Corporation recalls Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-DxThere is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-DxThere is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-DxThere is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power CordA subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
-
Class IIofficial
Boston Scientific Corporation recalls Suture Cinch Long 1pk. Endoscopic tissue approximation device.Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.
-
Class IIofficial
Boston Scientific Corporation recalls OverStitch Suture CinchDevice may fail to deploy the cinch implant and/or fail to cut the suture as intended.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific ENROUTE Transcarotid Neuroprotection SystemPotential for arterial sheath tip separation or partial tip separation during use.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific ENROUTE Transcarotid Neuroprotection System PlusPotential for arterial sheath tip separation or partial tip separation during use.
-
Class IIofficial
Boston Scientific Corporation recalls Radial Jaw 4 Standard Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 SC 240cPotential for breach in sterile barrier of device packaging.
-
Class IIofficial
Boston Scientific Corporation recalls Radial Jaw 4 Large Capacity Biopsy Forceps with the following descriptions and UPN: 1. Radial Jaw 4 LC 240cm BPotential for breach in sterile barrier of device packaging.
-
Class IIofficial
Boston Scientific Corporation recalls Radial Jaw 4 Jumbo Biopsy Forceps with the following descriptions and UPN: Radial Jaw 4 Jumbo 240cm Box 5; OutPotential for breach in sterile barrier of device packaging.
-
Class IIofficial
Boston Scientific Corporation recalls Rapid Refill Continuous Injection SystemDevices contain recalled Medline Namic Angiographic Control Syringe with Rotating Adaptor. There is a risk of syringe loosening or disconnecting from the manifold, resulting in loose connection or complete disconnection. May lead to difficulty connecting syringe, inability to aspirate or inject properly, fluid leakage, or separation of the syringe from the system, potentially requiring device exchange and prolonging the procedure.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific TheraSphere 360 Y-90 Management PlatformSoftware anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.
-
Class IIofficial
Boston Scientific Corporation recalls FARADRIVE STEERABLE SHEATH (CLEAR) - USBoston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
-
Class IIofficial
Boston Scientific Corporation recalls FARADRIVE Steerable Sheath Clear IDEBoston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
-
Class IIofficial
Boston Scientific Corporation recalls FARADRIVE STEERABLE SHEATH CLEAR OUSBoston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
-
Class IIofficial
Boston Scientific Corporation recalls FARASTAR PFA GeneratorLabeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
-
Class IIofficial
Boston Scientific Corporation recalls FARASTAR Pulsed Field Ablation GeneratorLabeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
-
Class IIofficial
Boston Scientific Corporation recalls FARASTAR Ablation Generator EULabeling discrepancy: Devices distributed with Instructions For Use (IFU) did not include FARAPOINT PFA catheter information.
-
Class IIofficial
Boston Scientific Corporation recalls FARASTAR Ablation Generator EUBoston Scientific is replacing specific FARASTAR PFA generators which have a configuration that is not compatible for use with the FARAPOINT PFA catheter.
-
Class IIofficial
Boston Scientific Corporation recalls FARADRIVE STEERABLE SHEATH CLEARBoston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Pro Wireguided 18-20mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Wireguided 10-12mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Pro Wireguided 8-10mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Wireguided 18-20mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Wireguided 6-8mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Wireguided 15-18mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Wireguided 12-15mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Wireguided 8-10mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Pro Wireguided 6-8mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Pro Wireguided 15-18mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Pro Wireguided 10-12mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls CRE Pro Wireguided 12-15mm 240cmPotential sterile breach of the pouches in which devices are packaged.
-
Class IIofficial
Boston Scientific Corporation recalls Orca Single Use Air/Water and Suction ValvesBoston Scientific is initiating a medical device removal of Orca Sterile, Single Use Air/Water and Suction Valves due to a higher than anticipated occurrence of the suction button sticking.
-
Class IIofficial
Boston Scientific Corporation recalls Intera Oncology INTERA 3000 Hepatic Artery Infusion PumpThere is the potential for leakage along the drug pathway from the pump through the end of the catheter.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-Dx II PlusFor some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-DX IIFor some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific PacemakerUpdated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P ELUpdated software, Brady software maintenance release 6 (SMR6), is now available for the ACCOLADE" family of pacemakers and cardiac resynchronization therapy pacemakers (CRT-Ps)1 and the advisory population is expanding to include all CRT-P and dual-chamber extended life (DR-EL) devices. Dual chamber modes are specifically indicated for treatment of the following: " Conduction disorders that require restoration of AV synchrony, including varying degrees of AV block " VVI intolerance (i.e., pacemaker syndrome) in the presence of persistent sinus rhythm " Low cardiac output or congestive heart failure secondary to bradycardia
-
Class IIofficial
Boston Scientific Corporation recalls EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M0050900Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.
-
Class Iofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery-Enhanced Delivery System - 8mm x 8mm; Material Number (UPN): M00553690;Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
-
Class Iofficial
Boston Scientific Corporation recalls *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 8mm x 8mm; Material Number (UPN): M00553530;Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
-
Class Iofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery-Enhanced Delivery System - 6mm x 8mm; Material Number (UPN): M00553680;Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
-
Class Iofficial
Boston Scientific Corporation recalls *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x 10mm; Material Number (UPN): M00553560;Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
-
Class Iofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
-
Class Iofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation CatheterBoston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.
-
Class IIofficial
Boston Scientific Corporation recalls CX Preconnect with TENACIO Pump with InhibiZone; UPNs: (1) UPN: 72404430 (12CM PS IZ)The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls LGX Preconnect with TENACIO Pump with InhibiZone; UPNs: (1) UPN: 72404450 (12CM PS IZ)The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls CX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404486 (15CM IP)The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls CXR Preconnect with TENACIO Pump with InhibiZone; UPNs: (1) UPN: 72404461 (12CM PS IZ)The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls LGX Preconnect with TENACIO Pump without InhibiZone; UPNs: (1) 72404405 (12CM IP)The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls TENACIO Pump with InhibiZone; UPN: 72404429;The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls TENACIO Pump without InhibiZone; UPN: 72404420;The potential for devices to experience inflation and/or deflation performance issues or difficulties.
-
Class IIofficial
Boston Scientific Corporation recalls Extractor Pro RX Retrieval Balloon Catheter; UPN (REF): (1) M00547030 (9-12mm Below)The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE DR SL MRI (Model L311)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number S722Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls PROPONENT DR SL (Model L201)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE SR SL MRI (Model L310)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L100Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number U125Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE SR SL (Model L300)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L111Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number S701 ALTRUA 2 SR SL PacemakerSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE DR EL (Model L321)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE DR SL (Model L301)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L210 PROPONENT SR SL PacemakerSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number U225 VISIONIST CRT-P ELSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L110Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L121 ESSENTIO DR EL PacemakerSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE DR EL MRI (Model L331)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L131Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L221 PROPONENT DR EL PacemakerSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number U228Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls PROPONENT SR SL (Model L200)Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number U226Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L209 PROPONENT DR (VDD) SL PacemakerSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L101Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L231Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number U128Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number S702Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class Iofficial
Boston Scientific Corporation recalls Model Number L211 PROPONENT DR SL MRI PacemakerSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
-
Class IIofficial
Boston Scientific Corporation recalls Preventive Maintenance (PM) Kit ICEFXCertain desiccant tube subassemblies were built with end caps that were not correctly tightened.
-
Class IIofficial
Boston Scientific Corporation recalls ICEfx Cryoablation SystemCertain desiccant tube subassemblies were built with end caps that were not correctly tightened.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific Encore Advantage KitThe Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific Encore 26 Inflation DeviceThe Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific UroMax Ultra KitThe Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific NephroMax KitThe Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System ANTBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System ANTBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN FXD Curve Access System SglBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System SGLBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSteer Access SystemBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN FXD Curve Access Sys DblBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System DBLBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN FXD Curve Access Sys SglBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System DBLBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN FXD Curve Access System DblBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls WATCHMAN TruSeal Access System SGLBoston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
-
Class Iofficial
Boston Scientific Corporation recalls ENDOTAK RELIANCE G/SGAffected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
-
Class Iofficial
Boston Scientific Corporation recalls RELIANCE 4-FRONTAffected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
-
Class Iofficial
Boston Scientific Corporation recalls ENDOTAK RELIANCE G/SG with 4-SITE ConnectorAffected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
-
Class Iofficial
Boston Scientific Corporation recalls Carotid WALLSTENT Monorail Endoprosthesis Closed Cell Self-Expanding StentCertain batches are being removed due to an increase in complaints received in which physicians encountered greater than anticipated resistance while attempting to withdraw the stent delivery system (SDS) from the guidewire or embolic protection device (EPD) after successful stent deployment. The most serious potential adverse health consequence is stroke. Other risks include delay to procedure, vessel injury, vessel spasm, or stent disruption/damage necessitating additional intervention.
-
Class IIofficial
Boston Scientific Corporation recalls FARAWAVE 1.0 Pulsed Field Ablation CatheterThe potentially impacted units were manufactured using specific equipment that may have caused cracks in the electrode bands on the catheter's distal end.
-
Class IIofficial
Boston Scientific Corporation recalls VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSAPotential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
-
Class IIofficial
Boston Scientific Corporation recalls VersaCross RF Wire (VXW). Indicated for creation of an atrial septal defect in the heart. Product ID VXW0002 DPotential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
-
Class IIofficial
Boston Scientific Corporation recalls VersaCross Access Solution (VXSK)Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
-
Class IIofficial
Boston Scientific Corporation recalls Torflex Transseptal Guiding Sheath Kit (TFK-M) TF85-32-63-45 TORFLEX 8.5F 63CM 45DEG TF85-32-63-55 TORFLEX 8.5Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
-
Class IIofficial
Boston Scientific Corporation recalls IceSeed 1.5 CX NEEDLE USBoston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
-
Class IIofficial
Boston Scientific Corporation recalls IceSeed 1.5 CX S NEEDLE USBoston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
-
Class IIofficial
Boston Scientific Corporation recalls IceSeed 1.5 CX NEEDLE OUSBoston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
-
Class IIofficial
Boston Scientific Corporation recalls IceSeed 1.5 CX S NEEDLE OUSBoston Scientific has identified that IceSeedTM CX needles manufactured between April 2024 and February 2025 have been programmed with DEMO settings instead of Commercial settings (see Affected Product Listing). Upon initial connection of an IceSeed CX needle to the Cryoablation System, the user will be prompted to perform the Needle Integrity and Functionality Test (NIT) per standard protocols. The affected needles perform as intended unless they are disconnected and reconnected to any channel after testing. In this case, the Cryoablation System will prompt re-execution of the NIT.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553660Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Delivery System 15mmx10mm UPN: M00553750Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553520Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553570Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553530Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 8mmx8mm UPN: M00553690Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553640Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 6mmx8mm UPN: M00553680Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx15mm UPN: M00553670Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553650Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00553550Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 20mmx10mm UPN: M00553560Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 10mmx10mm UPN: M00553540Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class IIofficial
Boston Scientific Corporation recalls AXIOS Stent and Delivery System 10mmx10mm UPN: M00553740Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one.
-
Class Iofficial
Boston Scientific Corporation recalls ALTRUA 2 DR PacemakerA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
-
Class Iofficial
Boston Scientific Corporation recalls VISIONIST CRT-PA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
-
Class Iofficial
Boston Scientific Corporation recalls ESSENTIO DR PacemakerA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
-
Class Iofficial
Boston Scientific Corporation recalls ACCOLADE DR PacemakerA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
-
Class Iofficial
Boston Scientific Corporation recalls VALITUDE CRT-PA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
-
Class Iofficial
Boston Scientific Corporation recalls PROPONENT DR PacemakerA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
-
Class IIofficial
Boston Scientific Corporation recalls AMS Artificial Urinary Sphincter Pressure Regulating Balloon 61-70 cm H2OPotential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
-
Class IIofficial
Boston Scientific Corporation recalls AMS Artificial Urinary Sphincter Control PumpPotential for mislabeled packaging in which a device labeled as a Pressure Regulating Balloon (UPN 72400024) which instead contained a Control Pump (UPN 72400098)
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific POLARx BALLOON CATHETER ST 28MM OUSBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER ST OUSBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific SMARTFREEZE CRYO CONSOLE: Console: Material Number M004CRBS4000; CONSOLE REFURB: Material NuBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER LT OUSBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER LT USBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific POLARx FIT BALLOON CATHETER ST USBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class Iofficial
Boston Scientific Corporation recalls Boston Scientific POLARx BALLOON CATHETER LT 28MM OUSBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific myLUX Patient Application for an Apple Mobile devicePatients using the Model 2939 myLUX Patient Application on their Apple mobile device (e.g., iPhone) may not be transmitting data from their Insertable Cardiac Monitor to the LATITUDE Clarity website.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls Boston Scientific AVVIGO+ Multi-Modality Guidance SystemBoston Scientific has received reports indicating the grid mark overlay on the cross-sectional view did not present correctly when an OptiCross 18 Peripheral Imaging Catheter was connected to the AVVIGO+ Multi-Modality Guidance System while in either Live or Record mode. The Boston Scientific investigation determined that, due to a software anomaly, a 9-grid mark overlay is incorrectly displayed during Live mode when AVVIGO+ is connected to an OptiCross 18 Peripheral Imaging Catheter. Note that the correct overlay display for this catheter is the 15-grid mark overlay, which allows for a reference depth setting of 14.9mm for peripheral catheters. This software anomaly is limited to Live or Record mode. In Review mode, the correct 15-grid mark overlay is displayed, and the vessel image is correctly scaled to the appropriate size.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO FC RMV 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX PC ESOPH STENT 18/23MM X 103MM- Wallflex Esophageal Partially Covered and Fully Covered Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO OTW PC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV 23X125MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SyWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO OTW PC 23MM X 12.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO OTW FC 23MM X 15.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO OTW FC 23MM X 10.1CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV LL 23X105- Intended for maintaining esophageal luminal patency in esophageal strictuWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV LL 23X125-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SyWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO FC RMV 23MM X 15.0CM IN 18.5 F- Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW SWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX PC ESOPH STENT 23/28MM X 155MM- Wallflex Esophageal Partially Covered and Fully Covered Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV LL 23X155- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX ESOPHAGEAL FC 18/23-25MMX10CM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) StenWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX ESOPHAGEAL FC 18/23-25MMX12CM-- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) SteWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX ESOPHAGEAL FC 23/28MMX10CM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is IWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV 18X153MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SysWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO OTW PC 23MM X 15.0CM IN 18.5 F-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX PC ESOPH STENT 23/28MM X 125MM- Wallflex Esophageal Partially Covered and Fully Covered Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV 18X123MM-Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SysWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX PC ESOPH STENT 18/23MM X 153MM- Wallflex Esophageal Partially Covered and Fully Covered Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV 23X105MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SyWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO FC RMV 23MM X 6.2CM IN 18.5 FR-Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW StWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV 23X155MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SyWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV 18X103MM- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SyWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX PC ESOPH STENT 18/23MM X 123MM- Wallflex Esophageal Partially Covered and Fully Covered Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX ESOPHAGEAL FC 23/28MMX15CM-Wallflex Esophageal Partially Covered and Fully Covered Stent System is InWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX ESOPHAGEAL FC 18/23-25MMX15CM- Agile Esophageal Partially Covered and Fully Covered OTW Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX ESOPHAGEAL FC 23/28MMX12CM- Wallflex Esophageal Partially Covered and Fully Covered Stent System is IWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX PC ESOPH STENT 23/28MM X 105MM- Wallflex Esophageal Partially Covered and Fully Covered Stent SystemWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls AGILE ESO OTW PC 23MM X 6.2CM IN 18.5 FR- Agile Esophageal Partially Covered (PC) and Fully Covered (FC) OTW SWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls WALLFLEX FC ESO STENT RMV LL 18X103- Wallflex Esophageal Partially Covered (PC) and Fully Covered (FC) Stent SWallFlexTM Esophageal Stent System and AgileTM Esophageal Over the Wire (OTW) Stent System due to the potential for delivery catheter tip detachment.
-
Class IIofficial
Boston Scientific Corporation recalls Flexima APDL Drainage Catheter System and KitThere is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.
-
Class IIofficial
Boston Scientific Corporation recalls Flexima Regular Kit Biliary Catheter System KitThere is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.
-
Class IIofficial
Boston Scientific Corporation recalls vanSonnenberg Sump Locking Pigtail Regular Kit Sump Catheter System KitThere is the potential for holes in the clear sterile barrier of the device pouch which may impact device sterility.
Search every recall that mentions Boston Scientific Corporation
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Boston Scientific Corporation or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts