Ventec Life Systems, Inc. recalls V+Pro Model Number: PRT-01185-000
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
- Products
-
- V+Pro Model Number: PRT-01185-000
- Company
- Ventec Life Systems, Inc.
- Where it was sold
-
Distribution US nationwide including Puerto Rico.
Nationwide
- Barcodes
008555730078770085001876116100850018761154- Units affected
- 39,568 units
- About this recall
- V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.