Medtronic Neuromodulation recalls Vanta Implantable Neurostimulator (INS)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Model 977006 Vanta INS batterty life has depleted faster than HCPs or patients expected leading to outpatient reprogramming or earlier than anticipated surgical replacements.
- Products
-
- Vanta Implantable Neurostimulator (INS)
- Company
- Medtronic Neuromodulation
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
0076300041192300763000478087007630005186150076300075424200763000849078- Units affected
- 11789 units
- About this recall
- Vanta Implantable Neurostimulator (INS), Model 977006