Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Insulet Corporation recalls The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
Products
  • The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
10385083000459
Units affected
174,013 units
About this recall
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.