Insulet Corporation recalls The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
- Products
-
- The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field
- Company
- Insulet Corporation
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
10385083000459- Units affected
- 174,013 units
- About this recall
- The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.