Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Insulet Corporation recalls Omnipod 5 Automated Insulin Delivery System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.
Products
  • Omnipod 5 Automated Insulin Delivery System
Where it was sold

US market only- Although there are 122823 downloads of the application, our data shows there are only ~80k active users.

Barcodes
10385083000336
Units affected
122,823 downloads
About this recall
Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product Description: The Omnipod 5 iOS App software is available to download through the Apple app store. The Omnipod 5 App is the user interface that controls the Omnipod 5 Automated Insulin Delivery System. It is used to activate/deactivate Pods, display alerts/alarms, and send insulin delivery commands for execution to the Pod. Only available in the US.