Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Shimadzu Corporation Medical Systems Division recalls Trinias unity

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
Products
  • Trinias unity
Where it was sold

US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.

States: Illinois, Kentucky, Massachusetts, Ohio, South Carolina, Tennessee

Barcodes
04540217058327
Units affected
13
About this recall