Shimadzu Corporation Medical Systems Division recalls Trinias unity
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
- Products
-
- Trinias unity
- Where it was sold
-
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.
States: Illinois, Kentucky, Massachusetts, Ohio, South Carolina, Tennessee
- Barcodes
04540217058327- Units affected
- 13
- About this recall
- Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the following main components: C arm, catherization table, image processing equipment, x-ray high voltage generator, x-ray tube. There are several types of each component.