FDAMedical devicesRecalled Status: Ongoing

Shimadzu Corporation Medical Systems Division recalls SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscop

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
Products
  • SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscop
Where it was sold

US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.

States: Arkansas, California, Colorado, Connecticut, Florida, Hawaii, Iowa, Illinois, Louisiana, Massachusetts, Maryland, Minnesota, Montana, North Carolina, Nebraska, New York, Ohio, Tennessee, Texas, Washington, West Virginia

Barcodes
04540217064809
Units affected
3
About this recall
SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscopic examinations, including general radiography and pediatric examinations, excluding mammography.