Shimadzu Corporation Medical Systems Division recalls SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscop
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography mode, which may lead to the possibility of delay or cancellation of the examination.
- Products
-
- SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscop
- Where it was sold
-
US Nationwide distribution in the states of CA, CT, MA, WV, OH, NY, IA, NE, MN, IL, FL, NC, HI, LA, MD, MT, WA, AR, TX, TN, CO.
States: Arkansas, California, Colorado, Connecticut, Florida, Hawaii, Iowa, Illinois, Louisiana, Massachusetts, Maryland, Minnesota, Montana, North Carolina, Nebraska, New York, Ohio, Tennessee, Texas, Washington, West Virginia
- Barcodes
04540217064809- Units affected
- 3
- About this recall
- SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscopic examinations, including general radiography and pediatric examinations, excluding mammography.