Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips Ultrasound, LLC recalls Philips Lumify Diagnostic Ultrasound System

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
Products
  • Philips Lumify Diagnostic Ultrasound System
Where it was sold

Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.

Nationwide

Units affected
13,206
About this recall
Philips Lumify Diagnostic Ultrasound System