Philips Ultrasound, LLC recalls Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.
- Products
-
- Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234
- Company
- Philips Ultrasound, LLC
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam.
States: Alabama, California, Connecticut, Florida, Georgia, Kansas, Maryland, Minnesota, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin
- Units affected
- 219 units
- About this recall
- Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234