American Contract Systems Inc. recalls Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- Products
-
- Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC
- Where it was sold
-
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
States: Iowa, Illinois, Minnesota, South Dakota, Washington
- Barcodes
191072229366191072207357191072211477191072236333191072203519191072226303191072226372191072236500191072236531191072235930191072223029191072214072191072235565- Units affected
- 32,433 total
- About this recall
- Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL PACK, MHLP45AJ LAPAROSCOPY PACK - 206005, MHRB88AO ROBOTIC PACK - 242561, RCLC51C LAP CHOLE RCH 0246969, SMLC53C LAP CHOLE PACK, UDLC17AE DSC LAP CHOLE PACK - 206062, UIRU84AE ROBOTIC UROLOGY PACK (PS 060571)