American Contract Systems Inc. recalls Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Products
  • Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO
Where it was sold

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States: Iowa, Illinois, Minnesota, South Dakota, Washington

Barcodes
19107223535019107222821519107215992219107222886419107214141500191072213990191072205537
Units affected
32,433 total
About this recall