Olympus Corporation of the Americas recalls Olympus Gastrointestinal Videoscope
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
- Products
-
- Olympus Gastrointestinal Videoscope
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
04953170420191- Units affected
- 1407 units
- About this recall
- Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100