Olympus Corporation of the Americas recalls
Olympus Corporation of the Americas is named on 78 recalls in Recall Relay's records since November 2023, from FDA (254 recalled products: FDA lists each product as its own recall); the most recent was August 27, 2026.
The newest 200 of 254 recalled products, newest first.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Tripp-Lite P580-025 25ft DisplayPort CableWhile in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus M3 Gold Autoclavable Lateral Telescope 70. Model Number: M3-70A.There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus BiCOAG Hemostasis Probe 10Fr Fixed Pin Connector. Model/Catalog Number: CD-B612LA.Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus BiCOAG Hemostasis Probe 7Fr Coaxial Connector. Model/Catalog Number: CD-B620LA.Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus BiCOAG Hemostasis Probe 10Fr Coaxial Connector. Model/Catalog Number: CD-B610LA.Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus BiCOAG Hemostasis Probe 7Fr Fixed Pin Connector. Model/Catalog Number: CD-B622LA.Potential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
-
Class IIofficial
Olympus Corporation of the Americas recalls POWERSEAL 5 MM (in diameter) 23 CM (long)Potential labeling deficiency in which an outer shelf box is incorrectly labeled as a 37-cm device.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for dPotential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Single Use Distal CoverOlympus has developed and released newly designed version of the Distal Cover (MAJ-2315) to replace previous design of the Single Use Distal Covers.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Ellik Evacuator Includes Glass BodySome sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2330.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2111Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the OlyPotential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2112Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2138.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2114.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2119.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.Potential for the Version 2 reprocessor connecting tube lock levers may fail prematurely.
-
Class IIofficial
Olympus Corporation of the Americas recalls OER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.XXX
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic ModePotential for detachment of a distal tip component of the device during use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic ModePotential for detachment of a distal tip component of the device during use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic ModePotential for detachment of a distal tip component of the device during use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat II Shears w/Ultrasonic ModePotential for detachment of a distal tip component of the device during use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus SOLTIVE Premium SuperPulsed Laser System Product Name: SOLTIVE Premium Laser System TFL-PLFollowing a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus HALO PKS Cutting ForcepsOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus PKS Cutting ForcepsOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Everest Bipolar Cutting Forceps w/CordOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Everest Bipolar Cutting ForcepsOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus PK Cutting ForcepsOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus PKS Cutting Forceps w/CordOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus PKS Cutting ForcepsOlympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Endoscope Air/Water Valve - MAJ-1444. Model Number: MAJ-1444.The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Endoscope Reprocessor OER-Elite. Model Number: N5789340.The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Endoscope Reprocessor OER-Pro. Model Number: N3058140.The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Endoscope Suction Valve - MAJ-1443. Model Number: MAJ-1443.The MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OER-Pro and OER-Elite automated endoscope reprocessors.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Single Use Biopsy Valve. Model/Catalog Number: MAJ-1218. 20 units per box. The MAJ-1218 has been desigPotential for rubber fragment detachment during use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designePotential for rubber fragment detachment during use.
-
Class Iofficial
Olympus Corporation of the Americas recalls Olympus High Flow Insufflation Unit. Model Number: UHI. Used for insufflation of the abdominal cavityIssue with software algorithm which may lead to overpressure events.
-
Class Iofficial
Olympus Corporation of the Americas recalls Olympus High Flow Insufflation Unit. Model/ Number: UHI-2. Used for insufflation of the abdominal cavityIssue with software algorithm which may lead to overpressure events.
-
Class Iofficial
Olympus Corporation of the Americas recalls Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavityIssue with software algorithm which may lead to overpressure events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy SyMis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC433Q-0720; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0720; Software Version: N/A PDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0725 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1530 Product Description: TheDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0720 Product DeDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC412Q-0215; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Disposable Triple Lumen Sphincterotome Product Name: Disposable Triple Lumen Sphincterotome Model/Devices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0330; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0730; Software Version: N/A PDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3020 Product Description: TheDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3030 Product Description: TheDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0730; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0320; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0730; Software Version: N/A PDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number:KD-VC433Q-0730; SoDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: CleverCut Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0320; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0720 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC431Q-0730; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V431M-0720; Software Version: N/A PDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0720; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-411Q-0730 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0725; SDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0720 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-431Q-0720 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-3025 Product Description: TheDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0730 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0725; Software Version: N/A PDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-0330; Software Version: N/A PDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: TDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: TheDevices which did not undergo thermoforming could deform and lose performance.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single UseOngoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: ShockPulse-SE Lithotripsy System SPL-SR Product Name: ShockPulse-SE Lithotripsy System - ReusableOngoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: ShockPulse-SE Lithotripsy System Product Name: ShockPulse-SE Lithotripsy System - Generator Model/Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner Sheath. Model No. A2642. for endoscopic diagnosis and treatment in urological applications.Complaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection Sheath. Model No. A22014A and A22014T. for endoscopic diagnosis and treatment in urologicalComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner Sheath 21 Fr. Model No. A2660T. for endoscopic diagnosis and treatment in urological applicationComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection Sheath. Model No. A2666T. for endoscopic diagnosis and treatment in urological applications.Complaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection Inner Sheath for 26 Fr. Outer Sheath. Model No. A22040T. Resection sheath for urologic appliComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner Sheath. Model No. A4741. for endoscopic diagnosis and treatment in gynecological applications.Complaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection Sheath; Model No. A2666. Resection sheath for urologic applications.Complaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection Inner SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner Sheath; Model No. A2660. for endoscopic diagnosis and treatment in urological applications.Complaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner Sheath. Model No. A2641. for endoscopic diagnosis and treatment in urological applications.Complaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Inner SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Resection SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus SheathComplaints of the ceramic tip of the resection sheath breaking have been received.
-
Class Iofficial
Olympus Corporation of the Americas recalls Brand Name: Single Use Ligating Device Product Name: Olympus HX-400U-30 Model/Catalog Number: HX-400U-30 UsedPotential that the ligation loop was unable to release or detach as expected during use, causing the loop to become unintentionally fixed around patient anatomy.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat 5 mmFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat 5 mmFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat 5 mmFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat 5 mmFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus ThunderbeatFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Thunderbeat 5 mmFirm is initiating a removal due to continued reports of adverse events.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilaDevice kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Duodenovideoscope Product Name: Olympus TJF-Q190V Model/Catalog Number: TJF-Q190V PContinued reports of positive cultures and infections have identified updates to the reprocessing materials to minimize potential deviations in TJF duodenoscope reprocessing. Olympus is informing users of these changes, reminding users to closely follow the operation and reprocessing instructions, and providing updated materials to users.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: LithotriptorThe device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Single Use Electrosurgical Snare SD-400 Product Name: Single Use Electrosurgical Snare ModSterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Single Use Single Use Electrosurgical Snare SD-400 Product Name: Single Use Single Use EleSterility compromised from small holes that may occur during packaging of the device pouches into the shelf carton and/or during transportation of the shelf carton.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF-Q170Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE MP60Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1TQ180Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE P60Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS BF-XT190Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T150Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF-H1100Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOFIBERSCOPE OLYMPUS BF TYPE PE2Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF-1TQ170Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180-ACAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P150Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE 1T60Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls BRONCHOFIBERSCOPE OLYMPUS BF TYPE TE2Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls OES BRONCHOFIBERSCOPE OLYMPUS BF TYPE XT40Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE XT160Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
-
Class Iofficial
Olympus Corporation of the Americas recalls ViziShot 2FLEX. Single Use Aspiration Needle. Model Number: NA-U403SX-4019. The ViziShot 2 FLEX tissue biopsyPotential for undetected, deformed a-traumatic tips.
-
Class IIofficial
Olympus Corporation of the Americas recalls ElectrosurgicalPotential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
-
Class IIofficial
Olympus Corporation of the Americas recalls ElectrosurgicalPotential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
-
Class IIofficial
Olympus Corporation of the Americas recalls Blue SiliconeDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Green Silicone Seal for AccessoriesDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Ellik Evacuator AdapterDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Ellik Latex BulbDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Glass Body For 194 Ellik Evacuator. Model Number: 190A. The Ellik Evacuator and adapter accessories are compatDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Cone with Male TubingDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Tubing for Male Connector 194Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Ellik Evacuator. Model Number: 194. The Ellik Evacuator and adapter accessories are compatible with Gyrus ACMIDiscrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/CatalogPotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser Fibers Model/Catalog NumberPotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium SuperPulsed Laser System Model/CatPotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: Soltive SuperPulsed Laser Fibers - Single Use ModePotential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190VRecent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/CatalogPreset treatment parameters are not consistently being used in accordance with the IFU.
-
Class Iofficial
Olympus Corporation of the Americas recalls Olympus Single Use Guide Sheath Kits with the following product descriptions and model numbers: 1. Single UsePotential for the radiopaque tip of the guide sheath component to disassociate and fall off into the patient.
-
Class Iofficial
Olympus Corporation of the Americas recalls Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is desPotential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.
-
Class IIofficial
Olympus Corporation of the Americas recalls LithoCrushV Single Use Mechanical Lithotriptor V. Model Number: BML-V442QR-30. Single Use Mechanical LithotripPotential for distal tip tearing of the Mechanical Lithotriptor V.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: ThunderbeatThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat Front-ActThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: THUNDERBEAT 5 mmThe probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.
-
Class IIofficial
Olympus Corporation of the Americas recalls Duravent Silicone Ventilation Tube. Model Number: 240075.Potential that Duravent Silicone Ventilation Tubes may contain incorrect product, specifically, 1.32 mm Tiny T Tubes (model: 70240076) instead of the intended Duravent Silicone Ventilation Tubes.
-
Class IIofficial
Olympus Corporation of the Americas recalls EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190High-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
-
Class IIofficial
Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190 Product Name: Olympus GIF-1TH190 Model/CatalTo better define the drying procedures described in the IFU to improve the instructions for easier understanding of each drying step after human errors and concerns were identified when completing a Human Factors evaluation of the GIF-1TH190.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Cystoscope Outer Sheath; Model/Catalog number: WA22810A;Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device.
-
Class IIofficial
Olympus Corporation of the Americas recalls EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopicMislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/AOlympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.
-
Class IIofficial
Olympus Corporation of the Americas recalls Brand Name: UroPass Ureteral Access Sheaths Product Name: Ureteral Access Sheaths Model/Catalog Number: 61024BExposure of the UroPass product to Ultraviolet ( UV ) Radiation can cause brittleness of the device dilator tip, which may lead to broken dilator tips in the package or in patients during surgical procedures.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Electrosurgical GeneratorThere was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Endoscopic Flushing PumpDuring product testing, Olympus identified an intermittent loss of function of the OFP-2 Flushing Pump caused by an internal component connection failure.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Gastrointestinal VideoscopeIt was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive Pro SuperPulsed Laser Systems-Intended for incisionThe term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser - Intended for incisionThe term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
-
Class IIofficial
Olympus Corporation of the Americas recalls The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannulaFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
-
Class IIofficial
Olympus Corporation of the Americas recalls The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannulaFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
-
Class IIofficial
Olympus Corporation of the Americas recalls The Everest Bipolar Cutting Forceps are intended to be passed through a 5-mm or 10-mm cannulaFractures and breakages in packaging trays and Tyvek covers, which may result in a sterility breach
-
Class IIofficial
Olympus Corporation of the Americas recalls ESG-410 Electrosurgical Generator- In conjunction with electrosurgical accessories intended for cutting and coElectrosurgical Generators ESG-410 does not power on upon use, may result in delays in initiating treatment or prolonged surgery
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive Pro SuperPulsed Laser SystemDifficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive SuperPulsed Laser System Wireless FootswitchDifficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive Premium SuperPulsed Laser SystemDifficulties in pairing the wireless footswitch with the Soltive Laser, potentially delaying the surgical procedure.
-
Class IIofficial
Olympus Corporation of the Americas recalls EasySuite 4K Integrated Operating Room System Model/Catalog Number: EasySuite 4K- The EasySuite 4K system inclA subset of the ES4K systems do not have complete records for earth leakage testing. The earth leakage test is a series of measurements intended to confirm that the amount of electrical current flowing to the earth is within standard limits
-
Class IIofficial
Olympus Corporation of the Americas recalls OES 4000 HysteroscopeThe required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
-
Class IIofficial
Olympus Corporation of the Americas recalls InstaClear Lens Sheath. The InstaClear lens cleaner system and versatile sheath cleans and clears the lens ofPotential for distal end of the sheath to break off during a procedure.
-
Class IIofficial
Olympus Corporation of the Americas recalls POWERSEAL 5 mmUpon initial activation of the seal function, the POWERSEAL non-conformance results in an immediate Incomplete Seal Cycle tone with the accompanying message on the generator screen, and no energy would be delivered to the device resulting in delay in surgery.
Search every recall that mentions Olympus Corporation of the Americas
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Olympus Corporation of the Americas or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts