Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
- Products
-
- BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150
- Where it was sold
-
US Nationwide distribution
Nationwide
- Barcodes
495317030818549531703392644953170339196- Units affected
- 15942 devices
- About this recall
- BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160