Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Olympus Corporation of the Americas recalls BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Products
  • BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150
Where it was sold

US Nationwide distribution

Nationwide

Barcodes
495317030818549531703392644953170339196
Units affected
15942 devices
About this recall
BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Numbers BF-1T150, BF-1T60, BF-P150 BF-P60 and EVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Model Number XT160