Olympus Corporation of the Americas recalls EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
- Products
-
- EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190
- Where it was sold
-
US Nationwide distribution
Nationwide
- Barcodes
4953170335181495317033517449531703429124953170335198- Units affected
- 32211 devices
- About this recall
- EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS BF-Q170 Model Number BF-Q170