Olympus Corporation of the Americas recalls Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Does not meet the requirements for electromagnetic compatibility (EMC) of the standard IEC 60601-1-2. Electromagnetic interference occurring during a procedure may cause the device to malfunction requiring the replacement of the device. This could potentially result in a prolonged procedure and/or require additional medical intervention
- Products
-
- Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029
- Where it was sold
-
OH
States: Ohio
- Barcodes
08719324487047- Units affected
- 3 units
- About this recall
- Quest Spectrum 3 Fluorescent Imaging System Model/Catalog Number: ART-SYS-0029, ART-SYS-0030 The Quest Spectrum3 is a fluorescent imaging system used in capturing and viewing fluorescent images for the visual assessment of blood flow for the evaluation of tissue perfusion, and related tissue-transfer circulation in tissue and free flaps used during plastic and reconstructive surgeries.