Philips Medical Systems Nederland B.v. recalls Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
- Products
-
- Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
- Where it was sold
-
Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.
States: Missouri, Texas
- Barcodes
00884838091535- Units affected
- 29 units
- About this recall
- Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system