Philips Medical Systems Nederland B.V. recalls
Philips Medical Systems Nederland B.V. is named on 52 recalls in Recall Relay's records since November 2023, from FDA (204 recalled products: FDA lists each product as its own recall); the most recent was October 6, 2026.
The newest 200 of 204 recalled products, newest first.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 3 M15. Model Number: 722064Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7 B12. Model Numbers: 722067Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7 M20. Model Numbers: 722079Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 3 M12. Model Numbers: 722063Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7 M12. Model Numbers: 722078Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7 B20. Model Numbers: 722068Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion S. Model Number: 722285.Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 5 M20. Model Numbers: 722228Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 5 M12. Model Number: 722227.Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion RystemTwo software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips AzurionTwo software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD. Model Numbers: 722026Firmware issue potentially leading to loss of geometry movements.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; MPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; ModPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips AlluraPotential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 forPhilips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model #Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001)The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (7220The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descrPotential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063)The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) inPotential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Vue Motion V12. Product Number: 1017979.Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD10; System Code: (1) 722003Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B12; System Code: (1)722067Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10C; System Code: 722001;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20; System Code: (1)722079Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD20; System Code: (1) 722006Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20; System Code: 722038;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15; System Code: 722058;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD10 OR Table; System Code: (1) 722022Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD10/10; System Code: (1) 722005Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20 OR Table; System Code: 722039;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M20; System Code: (1)722228Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M12; System Code: (1)722078Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M15; System Code: (1)722064Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD20 OR Table; System Code: (1) 722015Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/10; System Code: 722029;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD10F; System Code: 722002;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M12; System Code: (1)722227Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B20; System Code: (1)722068Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M12; System Code: (1)722063Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15 OR Table; System Code: 722059;Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls ALLURA Xper FD20 Biplane; System Code: (1) 722008Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M20 System Model Numbers: (1)722228Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B20 System Model Numbers: (1)722068Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M12 System Model Numbers: (1) 722063Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M15 System Model Numbers: (1) 722064Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M12 System Model Numbers: (1)722227Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B12 System Model Numbers:(1) 722067Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20 System Model Numbers: (1)722079Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M12 System Model Numbers: (1)722078Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20; Model Number: 722006;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10; Model Number: 722026;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane; Model Number: 722013;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 OR Table; Model Number: 722023;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane; Model Number: 722008;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10; Model Number: 722010;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/10; Model Number: 722029;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20; Model Number: 722012;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10/10; Model Number: 722005;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20; Model Number: 722038;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10/10; Model Number: 722011;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10; Model Number: 722003;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Product Name: Allura Xper FD10C; Model number: 722001;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20; Model Number: 722028;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10/10; Model Number: 722027;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10F; Model Number: 722002;Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptionsSix (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion system; Software Version Number: R3.1;Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray systemOver time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/10; Product Code: 722029; Associated 510(k) numbers: K102005Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Product Codes: (1)722012Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10/10; Model Numbers: (1) 722005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20; Model Numbers: (1) 722006The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane OR Table Product Codes: (1)722020Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 OR Table; Model Numbers: (1) 722015The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Cardio Vascular-Allura Centron; Model Numbers: 722400; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20; Model Numbers: 722038; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane; Model Numbers: (1) 722008The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 OR Table Product Codes: (1)722023Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper CV20; Model Numbers: 722031; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane OR Table; Model Numbers: (1) 722020The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10 Product Codes: (1)722010Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10C; Model Number: 722001; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10 OR Table; Model Numbers: 722033; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10; Model Numbers: (1) 722003The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10 OR Table Product Codes: (1)722022Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10/10 Product Codes: (1)722011Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane Product Codes: (1)722013Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
-
Class IIofficial
Philips Medical Systems Nederland B.V. recalls IntelliSpace CardiovascularSoftware issue that results in the display of outdated information.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponPotential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura System configured with a 1 Phase UPS. Allura family (R1.xPhilips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.xPhilips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura R8.2.x Systems. Software Verion: R8.2.x. 1. Allura Xper FD10Potential for temporary loss of imaging (X-ray) functionality due to software issue.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20. Fluoroscopic X-Ray System.Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion R1.x and R2.x systems; Model Numbers: (1) 722063A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
-
Class IIofficial
Philips Medical Systems Nederland B.V. recalls IntelliSpace Cardiovascular software versions 6.xAn issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera CVThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.0T Power/PulsarThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.5TThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.5T R11The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.5T Power/PulsarThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 3.0T Quasar DualThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 0.5T StandardThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.5T Omni/StellarThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Enterprise 1.5TThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.0T Omni/StellarThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls SmartPath to dStream for 1.5TThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.5T Explorer/Nova DualThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls SmartPath to dStream for 3.0TThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Intera 1.5T Master/NovaThe Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
-
Class IIofficial
Philips Medical Systems Nederland B.V. recalls Spectral CT on RailsMultiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10/10; Catalog numbers: (1) 722005Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M20; Catalog numbers: (1) 722228Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20; Catalog numbers: 722038Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10C System Code: (1) 722001Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXper FD20 Biplane OR Table System Code: (1) 722020 (2) 722025Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Cardio Vascular-Allura Centron; Catalog number: 722400 (OUS ONLY).Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10C; Catalog number: 722001Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10 - Catalog numbers: (1) 722003Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/10; Catalog numbers: 722029Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M12; Catalog numbers: (1) 722078Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B12; Catalog numbers: (1) 722067Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M12 System Code: (1) 722078 (2) 722223 (3) 722233Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/20 System Code: (1) 722038Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M20 System Code: (1) 722228 (2) 722232 (3) 722281Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B20; Catalog numbers: (1) 722068Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 OR Table; Catalog numbers: (1) 722023Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M12 System Code: (1) 722227 (2) 722231Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXperFD20/15 System Code: (1) 722058Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXper FD20 OR Table System Code: (1) 722023 (2) 722035Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 M12; Catalog numbers: (1) 722227Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3)Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20/15; Catalog numbers: 722058Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXperFD20/10 System Code: (1) 722029Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane OR Table; Catalog numbers: (1) 722020Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20; Catalog numbers: (1) 722006Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls AlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B20 System Code: (1) 722068 (2) 722226 (3) 722236Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Allura CV20 System Code: 722031Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M12; Catalog numbers: (1) 722063Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20; Catalog numbers: (1) 722079Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura CV20; Catalog numbers: 722031Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M15; Catalog numbers: (1) 722064Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class Iofficial
Philips Medical Systems Nederland B.v. recalls Allura Xper FD20 Biplane; Catalog numbers: (1) 722008Possibility of the patient falling from the table related to the mattress used on the patient table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 M12 System Code: (1) 722063 (2) 722221Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Zenition 70Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Zenition 50Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls System Name: 708038 MultiDiagnost Eleva with Flat Detector 708037 MultiDiagnost Eleva with Flat Detector 70803Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls System Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR TablPotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls System Name: 722044 INTEGRIS Allura 15-12 (biplane) 722030 Integris CV Cesar-Powerpack-Visub-Nicol 722021 INTEPotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 APotential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper systems (R8.1 with Poly-G Stand) Interventional Fluoroscopic X-Ray SystemA half-threaded bolt was included in the LTE kits instead of the correct full-threaded bolt.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted systemThe motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion Interventional Fluoroscopic X-ray System with Software Version Number: R1.0Potential Loss of Imaging (X-ray) Functionality and/or Longer Time to Perform Cold Restart
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722012When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper FD20 Biplane system with an ADN7NT patient table. Model Number: 722013When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper FD10 system with an ADN7NT patient table. Model Number: 722010When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Interventional Hemodynamic Application R.1.2XPotential for pressure wave data synchronization from two devices may not be correct.
-
Class IIofficial
Philips Medical Systems Nederland B.V. recalls Philips IntelliSpace Cardiovascular softwareStudy data is not able to be archived, copied, or exported with the cardiovascular software version.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 5 with a Certeray generator-To perform image guidance in diagnosticGenerator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 3 with a Certeray generator -To perform image guidance in diagnosticGenerator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnosticGenerator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper Interventional Fluoroscopic X-ray SystemsDual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper Interventional Fluoroscopic X-ray SystemsDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion Interventional Fluoroscopic X-ray SystemsDisk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Allura Xper Interventional Fluoroscopic X-ray SystemsFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion Interventional Fluoroscopic X-ray SystemsFramegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion systemSystem may exhibit a loss of imaging functionality and data due to an issue which can cause the system to continuously restart. If the issue occurs during a procedure, there may be a delay and/or abortion of the procedure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)Loss of connectivity between the FlexArm and the Table due to a software issue.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 5M12During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7B20/15During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 5M20During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M12During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7B12/12During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 3M12During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Search every recall that mentions Philips Medical Systems Nederland B.V.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Philips Medical Systems Nederland B.V. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts