Canon Medical System, USA, INC. recalls Cartesion Prime
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing), PET reconstruction control processing may hang up and PET reconstruction may stop as a result, which may necessitate repeat PET/CT examinations.
- Products
-
- Cartesion Prime
- Where it was sold
-
US: NV, AR, MN, IA, IL, NC, LA, KS, MS, CA, AL, MD, CO, IN, TN, AZ, WI, VA
States: Alabama, Arkansas, Arizona, California, Colorado, Iowa, Illinois, Indiana, Kansas, Louisiana, Maryland, Minnesota, Mississippi, North Carolina, Nevada, Tennessee, Virginia, Wisconsin
- Barcodes
04987670103810- Units affected
- 27
- About this recall
- Cartesion Prime, Digital PET-CT, Model: PCD-1000A/3D.004, PCD-1000A/3D, is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.