Canon Medical System, USA, INC. recalls ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
- Products
-
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V
- Where it was sold
-
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.
States: Alabama, Arkansas, Arizona, California, Florida, Georgia, Hawaii, Illinois, Louisiana, Massachusetts, Michigan, Minnesota, Mississippi, North Carolina, Nevada, Ohio, Oregon, Texas
- Units affected
- 31 units
- About this recall
- ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V