Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
Products
  • epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System
Where it was sold

US Nationwide. Canada.

Nationwide

Barcodes
00809708121860
Units affected
1292 units
About this recall
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515