Siemens Healthcare Diagnostics, Inc. recalls
Siemens Healthcare Diagnostics, Inc. is named on 43 recalls in Recall Relay's records since November 2023, from FDA (71 recalled products: FDA lists each product as its own recall); the most recent was August 17, 2026.
All 71 recalled products, newest first.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;The potential for falsely elevated results.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mLDue to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Beckman CoulterDue to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin REF: 11537225Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.
-
Class IIIofficial
Siemens Healthcare Diagnostics, Inc. recalls Syva Emit Disopyramide Assay Enzyme ImmunoassayOut-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
-
Class IIIofficial
Siemens Healthcare Diagnostics, Inc. recalls Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULout-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC)A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urine Albumin (UAlb). Material Number: 11537225Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls 3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1)The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP). Material Number: 11097543.Potential for falsely depressed patient, quality control (QC), and/or calibration results or delayed results.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA¿ 120/2120/2120i SETpoint CalibratorProducts have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN1 Control)Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (Norm Control)Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA¿ 120/2120/2120i 3-in-1 TESTpoint Control (ABN2 Control)Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc BGEM Crea Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736382;The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; eThe occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223.Incorrect software flagging may occur for the assay that may potentially lead to an erroneous result.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413583(CNSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413497 (EU); Software Version Number: 4.14.9. Intended for useSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413517(MX) ; Software Version Number: 4.14.9. Intended for useSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413506 (CA); Software Version Number: 4.14.9. Intended for useSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413475 (US); Software Version Number: 4.14.9. Intended for useSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413518 (ROW) ; Software Version Number: 4.14.9. Intended for uSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for useSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413498(JP) ; Software Version Number: 4.14.9. Intended for useSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Magnesium Flex reagent cartridgeSiemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CI Analyzer. Catalog Numbers: 10947347.Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Siemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reaction vA manufacturing defect causes Atellica CH Reaction Ring Cuvette Segments with lot numbers N15 on the packaging or 15 on the underside of the cuvette segment have a potential for imprecision and erroneously depressed or elevated QC and patient results for seven (7) analytes: Ammonia, Acetaminophen, Alanine Aminotransferase (ALT), Hemoglobin A1c, Creatinine Kinase (CK), Cystatin C, and Salicylate.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica¿ CH ¿2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of ¿2-microgloPotential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH ¿2-Microglobulin (B2M) reagent.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 PSA assayThe IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis SystemSiemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur aTPO AssayPositive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM aTPO AssayPositive bias in patient results at and below the cut-off of 60 U/mL with the lots listed in the table below when compared to alternate lots on the Atellica IM analyzer.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microaAtellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Microalbumin_2 (¿ALB_2)-For in vitro diagnostic use in the quantitative measurement of microalbumiAtellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11319151The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP); Siemens Material Number (SMN): 11097543The potential for biased quality control (QC) and patient results when using Atellica CH UCFP lot 130414 and ADVIA Chemistry UCFP lot 140 reagents.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM CA 19-9 (250 Test)- In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-aPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tuPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-aPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM CA 19-9 (50 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-assPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 5Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Anti-TG Ab Catalog # L2KTG6 (600 Tests)The potential for falsely elevated patient sample results when using impacted lots (785, 790, 791, and 793) of IMMULITE 2000 Anti-TG Ab.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 Anti-TG Ab Catalog # L2KTG2 (200 Tests)The potential for falsely elevated patient sample results when using impacted lots (782, 783, 785, 786, 787, 788, 789, 790, 791, 792 and 793), of IMMULITE 2000 Anti-TG Ab.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Iron3 - IVD use in the quantitative measurement of iron in human serum and plasma (lithium heparinPotential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positive bias ranging from 2-16% -impacts calibrator, quality control (QC), and patient results.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls DCA Systems HbA1c Reagent KitThe reason for the recall is that the Siemens DCA HbA1c Reagent Kit was confirmed to display a positive bias without alerting the user. The potential impact of the defect is that erroneous test results could be reported and potentially lead to incorrect patient management of glucose control. A positive bias may be observed with eleven HbA1c affected lots. Positive bias was observed when the performance of these five HbA1c lots was compared to NGSP pooled patient target-value assigned samples and the maximum bias observed was 19%.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur Erythropoietin (EPO) Assay (100 Test)Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Atellica IM Erythropoietin (EPO) Assay (100 Test)Siemens Healthcare Diagnostics has confirmed an average negative bias of -35% compared to the 3rd World Health Organization (WHO) International Standard (NIBSC code: 11/170) that is proportional across the assay measuring interval. The reference interval as claimed in the Instructions for Use (IFU) is not achieved with affected lots.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Total immunoglobulin M (IgM total) IVDPotential for a negative bias with quality control (QC) and patient sample results when using the Atellica CH Immunoglobulin M_2 (IgM_2) reagent.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 100 -Intended for in vitro diagnostic use and is dRAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 750-Intended for in vitro diagnostic use and is deRAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 400-Intended for in vitro diagnostic use and is deRAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPoint 500 Systems Measurement Cartridge (with Lactate) 250-Intended for in vitro diagnostic use and is deRAPIDPoint 500 Systems Measurement Cartridges (with lactate) has the potential to affect the Sodium (Na+) sensor to produce a negative bias that could result in serious injury due to a delayed diagnosis of hypernatremia or unnecessary intervention for hyponatremia and result in iatrogenic hypernatremia especially if the true sodium result is near the thresholds of severe alterations, as well as cause a Question Result -----? error flag for multiple electrolytes on patient samples and quality control that may cause a delay of diagnosis or treatment.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc NXS Host Blood Analysis SystemSiemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood Tests feature is enabled, the results that are generated at the time of testing are correct; however, when this issue occurs, results that are printed, saved and/or transferred to a data manager, may include unselected analytes that are from a previous patient or QC test. The worst possible outcome may result in a failure to diagnose hyperkalemia, due to an erroneous depressed potassium level or inappropriate treatment to hyperglycemia due to erroneous glucose reading.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPOINT 500e Blood Gas SystemThere are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls RAPIDPOINT 500 Blood Gas SystemThere are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.