Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ziehm-orthoscan Inc recalls Orthoscan VERSA Accessory Cart

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.
Products
  • Orthoscan VERSA Accessory Cart
Company
Ziehm-orthoscan Inc
Where it was sold

Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.

States: Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kansas, Massachusetts, Maryland, Michigan, Minnesota, North Carolina, Nebraska, Nevada, New York, Ohio, Oregon, South Carolina, Texas, Utah, Washington

Units affected
55 units
About this recall
Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020