Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
- Products
-
- VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
- Where it was sold
-
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
States: California, Colorado, Florida, Iowa, Illinois, Indiana, Kansas, Maryland, Missouri, Montana, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin
- Barcodes
04987670102615- Units affected
- 34 systems
- About this recall
- VANTAGE GALAN 3T Model MRT-3020/MEXL-3020