Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T Model MRT-3020/MEXL-3020

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Products
  • VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
Where it was sold

U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.

States: California, Colorado, Florida, Iowa, Illinois, Indiana, Kansas, Maryland, Missouri, Montana, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin

Barcodes
04987670102615
Units affected
34 systems
About this recall
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020