Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

FHC, Inc. recalls microTargeting Insertion Tube Set

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product with a spacer tube which is too long could result in intercranial hemorrhaging and brain damage.
Products
  • microTargeting Insertion Tube Set
Company
FHC, Inc.
Where it was sold

Domestic distribution only to the following states: AZ, CA, PA, TN, TX, WA.

States: Arizona, California, Pennsylvania, Tennessee, Texas, Washington

Units affected
8 packs of 5ea (40 total)
About this recall