GE Medical Systems, LLC recalls Senographe Pristina
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
- Products
-
- Senographe Pristina
- Company
- GE Medical Systems, LLC
- Where it was sold
-
USA and OUS.
- Barcodes
00840682118460- About this recall
- Senographe Pristina