GE Medical Systems, LLC recalls Optima XR200amx
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
- Products
-
- Optima XR200amx
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Nationwide
- Barcodes
00840682115278- Units affected
- 4 units
- About this recall
- Optima XR200amx, Mobile Digital-Ready Radiographic System