Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Products
  • VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Where it was sold

US distribution to Ohio and Nevada

States: Nevada, Ohio

Units affected
2 units
About this recall
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.