Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Wissner-bosserhoff Gmbh recalls SafeControl handset (a component of the Sentida bed which moves and positions the bed)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Products
  • SafeControl handset (a component of the Sentida bed which moves and positions the bed)
Company
Wissner-bosserhoff Gmbh
Where it was sold

US Nationwide distribution in the states of NC.

States: North Carolina

Barcodes
405062110040140506211000124050621100432405062110044940506211004704050621100487
Units affected
77 units
About this recall
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198