Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Baxter Healthcare Corporation recalls TruSystem 7000 U Operating Table (dV)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Products
  • TruSystem 7000 U Operating Table (dV)
Where it was sold

US Nationwide distribution in the states of IL, MO, TX.

States: Illinois, Missouri, Texas

Barcodes
00887761967519
Units affected
62 units
About this recall
TruSystem 7000 U Operating Table (dV), Product Code 1723633