Baxter Healthcare Corporation recalls TruSystem 7000 U Operating Table (dV)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
- Products
-
- TruSystem 7000 U Operating Table (dV)
- Where it was sold
-
US Nationwide distribution in the states of IL, MO, TX.
States: Illinois, Missouri, Texas
- Barcodes
00887761967519- Units affected
- 62 units
- About this recall
- TruSystem 7000 U Operating Table (dV), Product Code 1723633