Instrumentation Laboratory recalls GEM Premier 3000 PAK - pH
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
- Products
-
- GEM Premier 3000 PAK - pH
- Company
- Instrumentation Laboratory
- Where it was sold
-
US States: MA, MN, NY. India.
States: Massachusetts, Minnesota, New York
- Barcodes
084269500879220842695008250708426950082538- Units affected
- 2448 total
- About this recall
- GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004