FDAMedical devicesRecalled Status: Ongoing

Topcon Medical Systems, Inc. recalls The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided
Products
  • The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection
Company
Topcon Medical Systems, Inc.
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
04562123288582
Units affected
3144 units
About this recall
The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices. It is intended for processing and displaying ophthalmic images and optical coherence tomography data. The IMAGEnet 6 Ophthalmic Data System uses the same algorithms and reference databases from the original data capture device as a quantitative tool for the comparison of posterior ocular measurements to a database of known normal subjects.