Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Leoni Cia Cable Systems recalls ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.
Products
  • ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in
Where it was sold

US Nationwide distribution in the states of MA, TN.

States: Massachusetts, Tennessee

Barcodes
03700383510017
Units affected
1 unit
About this recall
ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00003