Leoni Cia Cable Systems recalls ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.
- Products
-
- ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in
- Company
- Leoni Cia Cable Systems
- Where it was sold
-
US Nationwide distribution in the states of MA, TN.
States: Massachusetts, Tennessee
- Barcodes
03700383510079- Units affected
- 3 units
- About this recall
- ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00009