GE Healthcare Finland Oy recalls Carescape B450
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
- Products
-
- Carescape B450
- Company
- GE Healthcare Finland Oy
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00840682146135008406821423800084068210522400840682142045- Units affected
- 3271
- About this recall
- Carescape B450, Model no. 5805686 - shipped with potentially affected batteries or affected batteries installed as a spare part Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms Record updated 3/19/26 to include: CARESCAPE B450 MBA313 2095800-001 CARESCAPE MONITOR B450 (ATO MODEL) 2068491-002 CARESCAPE MONITOR B450 B1 2107633-001 CARESCAPE MONITOR B450-GS (V2.0.8) 2094080-001