Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Healthcare Finland Oy recalls Portrait VSM Model nos. 6660000-666 Monitor

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
Products
  • Portrait VSM Model nos. 6660000-666 Monitor
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00195278561312
Units affected
5990
About this recall
Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms