GE Healthcare Finland Oy recalls Portrait VSM Model nos. 6660000-666 Monitor
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.
- Products
-
- Portrait VSM Model nos. 6660000-666 Monitor
- Company
- GE Healthcare Finland Oy
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00195278561312- Units affected
- 5990
- About this recall
- Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms