Sonesta Medical AB recalls Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Descr
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Video fluoroscopy table may contain inadequate weld on backrest bracket resulting in backrest dropping.
- Products
-
- Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Descr
- Company
- Sonesta Medical AB
- Where it was sold
-
US Distribution to NH only.
States: New Hampshire
- Units affected
- 10 tables
- About this recall
- Brand Name: Sonesta Product Name: 6210 Model/Catalog Number: 0120-LA120-03 Software Version: N/A Product Description: Sonesta 6210 - Procedure table designed for full fluoroscopic imaging of the urinary tract during video Urodynamics procedures. Component: N/A