Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Abbott Medical recalls Abbott Navitor Titan

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
Products
  • Abbott Navitor Titan
Where it was sold

New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.

States: New York

Barcodes
05415067045768
Units affected
1 US, 9 OUS
About this recall
Abbott Navitor Titan, Vision Technology, Transcatheter Aortic Heart Valve, 35 mm, NVRO-35, Sterile