Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.
Products
  • The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable
Where it was sold

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN TX, UT, VA, WA, WI, and WV. The countries of Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Nationwide

Barcodes
05414734509725
Units affected
7295 programmers
About this recall
The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.