Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

OrganOx Ltd recalls OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Nu

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Battery movement led to damage of one of the battery cells causing it to short to the battery retention bracket
Products
  • OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Nu
Company
OrganOx Ltd
Where it was sold

California, Texas, Ohio, Florida, New York, Massachusetts, Virginia, Oklahoma, DC, South Carolina and Georgia.

States: California, District of Columbia, Florida, Georgia, Massachusetts, New York, Ohio, Oklahoma, South Carolina, Texas, Virginia

Barcodes
05060462240029
Units affected
20 units
About this recall
OrganOx metra -Intended as a Normothermic Machine Perfusion system of donor transport livers. Model/Catalog Number: D0003 The OrganOx metra is a transportable device intended to be used to sustain donor livers destined for transplantation in a functioning state for a total preservation time of up to 12 hours.