Kreatech Biotechnology Bv recalls Kreatech PapV-16
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.
- Products
-
- Kreatech PapV-16
- Company
- Kreatech Biotechnology Bv
- Where it was sold
-
US States: FL, NY
States: Florida, New York
- Barcodes
08718858480241- Units affected
- 23 units
- About this recall
- Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495