Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Cellavision AB recalls REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.
Products
  • REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60
Company
Cellavision AB
Where it was sold

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.

Nationwide

Units affected
688 units
About this recall
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.